RPM vs RCF in Nucleic Acid Release: Recording Centrifugation Correctly

RPM and relative centrifugal force (RCF, expressed as x g) are different quantities. When transferring a nucleic-acid preparation procedure between centrifuges, record the rotor and the displayed unit. A speed written in rpm cannot be read as the same numerical value in x g.

Why the same rpm can mean a different force

RPM describes rotation rate. RCF also depends on the rotor radius. Using a radius in centimetres, the conversion is RCF = 1.118 x 10-5 x radius x rpm2. Use the rotor manufacturer’s radius definition and conversion guidance, including any relevant adapter.

Illustrative speed Illustrative radius Calculated RCF
12,000 rpm 6 cm Approximately 9,660 x g
12,000 rpm 8 cm Approximately 12,879 x g

These calculations illustrate the unit difference; they are not validated replacement settings for a DueBio reagent protocol.

What this means for BT0066 users

The BT0066 Lightning product instructions include a centrifugation step specified as 12,000 rpm for two minutes after heating and cooling. Preserve the stated units when documenting that procedure. If a receiving laboratory needs an equivalent force-based setting, first identify the rotor and clarify the intended transfer with technical support. Do not assume that 12,000 x g is interchangeable with 12,000 rpm.

Keep an equipment record alongside the reagent record

Record the centrifuge model, rotor identifier, tube and adapter, whether the display was set to rpm or x g, the specified duration and the source instruction version. Keep this information with the reagent lot and sample matrix. This makes a failed method transfer easier to investigate without first changing the chemistry.

A correct unit conversion alone does not establish nucleic-acid recovery or downstream amplification performance. Any protocol transfer still needs evaluation in its intended workflow. For evaluation batches or bulk/OEM discussions, contact DueBio/TiosBio with the reagent model, sample matrix and equipment details.

Sources and scope

Eppendorf: Basics in centrifugation explains the distinction between speed and force. Thermo Fisher: Centrifuge best practices provides the conversion formula. Product-specific handling remains governed by the applicable BT0066 instructions. No new DueBio performance data are reported here.

TL
Global Agent · Duebio (TiosBio) · 20+ Years in IVD
IVD industry veteran specializing in CRISPR Cas12/Cas13 detection, RAA isothermal amplification, lateral flow assays, microfluidic PCR, TRF immunoassays, and OEM/ODM IVD development for global distributors. Duebio is the international trade brand of TiosBio, a Chinese IVD manufacturer with 20+ years of experience.