RPM and relative centrifugal force (RCF, expressed as x g) are different quantities. When transferring a nucleic-acid preparation procedure between centrifuges, record the rotor and the displayed unit. A speed written in rpm cannot be read as the same numerical value in x g.
Why the same rpm can mean a different force
RPM describes rotation rate. RCF also depends on the rotor radius. Using a radius in centimetres, the conversion is RCF = 1.118 x 10-5 x radius x rpm2. Use the rotor manufacturer’s radius definition and conversion guidance, including any relevant adapter.
| Illustrative speed | Illustrative radius | Calculated RCF |
|---|---|---|
| 12,000 rpm | 6 cm | Approximately 9,660 x g |
| 12,000 rpm | 8 cm | Approximately 12,879 x g |
These calculations illustrate the unit difference; they are not validated replacement settings for a DueBio reagent protocol.
What this means for BT0066 users
The BT0066 Lightning product instructions include a centrifugation step specified as 12,000 rpm for two minutes after heating and cooling. Preserve the stated units when documenting that procedure. If a receiving laboratory needs an equivalent force-based setting, first identify the rotor and clarify the intended transfer with technical support. Do not assume that 12,000 x g is interchangeable with 12,000 rpm.
Keep an equipment record alongside the reagent record
Record the centrifuge model, rotor identifier, tube and adapter, whether the display was set to rpm or x g, the specified duration and the source instruction version. Keep this information with the reagent lot and sample matrix. This makes a failed method transfer easier to investigate without first changing the chemistry.
A correct unit conversion alone does not establish nucleic-acid recovery or downstream amplification performance. Any protocol transfer still needs evaluation in its intended workflow. For evaluation batches or bulk/OEM discussions, contact DueBio/TiosBio with the reagent model, sample matrix and equipment details.
Sources and scope
Eppendorf: Basics in centrifugation explains the distinction between speed and force. Thermo Fisher: Centrifuge best practices provides the conversion formula. Product-specific handling remains governed by the applicable BT0066 instructions. No new DueBio performance data are reported here.